Constructs a calibration model from standards with known concentrations.
Usage
ls_calibrate(
dataset,
element,
line_nm,
concentrations,
method = "univariate",
internal_std_nm = NULL,
window_nm = 1,
pls_window_nm = NULL,
n_components = 5,
unit = "ppm",
verbose = TRUE
)Arguments
- dataset
A
ls_dataset()containing standard spectra. Spectra order must correspond toconcentrations.- element
Character. Target element symbol.
- line_nm
Numeric. Emission line wavelength (nm).
- concentrations
Numeric vector of known concentrations, one per spectrum in
dataset.- method
Character.
"univariate"(default),"internal_std","pls", or"cf_libs".- internal_std_nm
Numeric. Internal standard line wavelength, required for
method = "internal_std".- window_nm
Numeric. Peak integration window. Default 1.
- pls_window_nm
Numeric vector of length 2. Spectral window for PLS. Default
c(line_nm - 5, line_nm + 5).- n_components
Integer. Number of PLS components. Default 5.
- unit
Character. Concentration unit. Default
"ppm".- verbose
Logical. Default
TRUE.
Value
A ls_calibration() object.
Examples
ds <- ls_example_data("calibration")
conc <- ds$sample_info$concentration
cal <- ls_calibrate(ds, "Ca", 393.37, conc, verbose = FALSE)
print(cal)
#> <libs_calibration>
#> • Element: "Ca"
#> • Line: 393.37 nm
#> • Method: "univariate"
#> • Unit: "ppm"
#> • Standards: 27
#> • R-squared: 0.9497
#> • LOD: 3.96 ppm
#> • LOQ: 13.2 ppm